Anesthesia machine AM-A6
Anesthesia machine AM-A6 ADVANCED DESIGN & MONITORING
Large screen high-definition TFT wide viewing angle color LCD touch screen
New interface design, reasonable layout and color matching
Pressure-Time waveform, Flow-Time waveform display on the same screen
Powerful monitoring and alarm functions
Comprehensive respiratory mechanics monitoring allows you to fully grasp the patient’s respiratory status
HIGH-PRECISION VAPORIZER
Vaporizer can support low flow anesthesia (as low as 0.2L/min and large flow anesthesia (up to 15L/min).
Accurate output of concentrations under varying conditions of pressure, flow and temperature.
Support Selectatec Mounting type and Cagemount compatible
4 Kinds of Vaporizer for Choice: Halothane, Enflurane, Isoflurane and Sevoflurane
Fill-in type: Pouring
INTEGRATED BREATHING CIRCUIT
Integrated breathing circuit with APL valve, excellent tightness and safety.
Made of special materials, resistant to corrosion and high temperature.
Compact structure, no external pipelines, good mechanical compliance.
Impeller sampling is used at the exhalation end to avoid inaccurate monitoring of tidal volume due to water vapor condensation.
Drive bellows is not required to repalce when operating pediatric anesthesia.
Large volume of CO2 canister, single tank structure for more convenient in assembly.
ACGO PORT
With ACGO port to connect with T circuit or BAIN circuit for open-operation.
Convenient for patients’oxygen inhalation and post-operative recovery.
AGSS PORT
AGSS exhaust gas collection system can be installed.
Reduce environmental pollution in the operating room Ensure the safety of medical staff and patients.
Anesthesia machine AM-A6 TECHNICAL PARAMETERS
PART OF ANESTHESIA MACHINE
■Main Body
High-strength engineering plastic rack, light, beautiful and corrosion resistant
■Scope of Application:
Adult, Child
■Gas Source
O2: 0.28~0.6MPa, N2O: 0.28~0.6MPa
■Flow Meter
O2: 0.05~1.0L/min,1.1~10L/min
N2O: 0.05~1.0L/min,1.1~10L/min
■O2–N2O Linkage and N2O Stopper
When using nitrous oxide, oxygen concentration>25%;
When the oxygen pressure<0.16±0.02Kpa, the flow of nitrous oxide would be cut off.
■Flow Rate of Rapid Oxygen Supply
25~75L/min
■Low Oxygen Pressure Alarm
There will be sound alarm when the oxygen pressure < 0.2MPa
■Vaporizer
It has the function of automatic compensation based on pressure, temperature, and flow rate. The regulation range of evaporator concentration is 0~5 vol%.
Among Halothane,Enflurane, Isoflurane and Sevoflurane, two can be chosen for application as required by the customer.
■Respiratory Circuit
Working mode: All-close, Semi-close, Semi-open
APL: ≤12.5 kPa
■Respiratory Bellows
Bellows for adults and children
Tidal volume range:0~1500 ml
■Light
Equipped with LED illumination, easy for observation
■ ACGO Port
With ACGO port to connect with T circuit or BAIN circuit for open-operation.
■ AGSS Port
AGSS exhaust gas collection system can be installed.
Anesthesia machine AM-A6 PART OF VENTILATOR
■Display Mode
High-definition 10.1”TFT color LCD touch screen display, easy and convenient operation
■Ventilation Mode
IPPV, SIPPV, SIMV, MANUAL
■Ventilation Function
PEEP,SIGH, IRV
■Ventilation Parameters
Tidal volume:50~1500 ml
Rate:2~99 bpm
SIMV rate:2~20 bpm
I:E:2:1~1:8
Inspiratory trigger pressure: -10~20 cmH₂O
PEEP:2~20 cmH₂O
Pressure range:5~60 cmH₂O
SIGH:1.5 times the inspiratory time/ 100 times breath
■Anesthesia machine AM-A6 Parameters For Ventilation Monitoring
Tidal Volume, Ventilation Volume, IPPV Rate, SIMV Rate, I:E, Peak Pressure of Airway, Pressure-Time Waveform, Flow Rate-Time Waveform, PEEP, Inspiratory Trigger Pressure (Average Pressure, FiO2 Oxygen Concentration, Battery Level)
■Security Alarm System
Oxygen Concentration Alarm: Upper limit setting range:21%~99%,Low limit setting range:18%~80%
Airway Pressure Alarm: Upper limit setting range:5~60 cmH₂O,Low limit setting range:0~1250pxH₂O
Per-minute Ventilation Volume Alarm: Upper limit setting range:1~20L/min,Low limit setting range: 0~19L/min
Tidal Volume alarm: Upper limit setting range:100~1500ml,Low limit setting range 0~1400ml
Sustained High Pressure Alarm: It will give alarm when stress have consistently been higher than 25 cmH₂O for 15s.
Suffocation Alarm: It will give alarm when stress have consistently been higher than 10 cmH₂O and tidal volume is less than 120ml for 15s.
Tube Escape Alarm: It will give alarm when stress have consistently been less than 50pxH₂O for 15s.
Low Battery Level Alarm: It will give alarm when battery level is less than 14±0.5V
Negative Pressure Alarm: It will give alarm when airway stress has been less than 250pxH₂O for 1s.
■Power Supply
AC 220V 50Hz
Built-in Li Battery (Work at least for 8 hours)
■ Options
Mechanical Arms Bracket for Patient Monitor
ETCO2 monitoring